2024-11-21 Version 2.3.3
Simple Search | Advanced Search | Pre-Defined Reports | Help

Details of 'REN/ERM-TG30-002-1' Work Item
Work Item Reference ETSI Doc.
Number
STF Technical Body
in Charge
Download Standard

Download standrd ref:
REN/ERM-TG30-002-1 (PDF Format) Download standrd ref:
REN/ERM-TG30-002-1 (Word Format)
  REN/ERM-TG30-002-1 EN 301 839-1   ERM TG30
  Current Status
(Click to View Full Schedule)
Latest
Version
Cover Date Standstill Creation Date
  Publication (2007-07-23) 1.2.1 2007-07-23 View Standstill Information 2004-07-01
  Rapporteur Technical Officer Harmonised Standard
  Joachim Elsner Marcello Pagnozzi No View Transposition Data
 
Title Electromagnetic compatibility and Radio spectrum Matters (ERM); Short Range Devices (SRD); Ultra Low Power Active Medical Implants (ULP-AMI) and Peripherals (ULP-AMI-P) operating in the frequency range 402 MHz to 405 MHz; Part 1: Technical characteristics and test methods
ULP AMI in the band 402 to 405 MHz 
Scope and Field
of Application
EN 301 839 shall be revised:
1. Generic definitions shall be used and inconsistencies be removed in EN 301 839-1 and -2.
2. At present EN 301 839-1 describes LBT as interference mitigation technology. It is suggested that also an alternative interference mitigation technology is introduced in EN 301 839-1. Possible candidates for this interference mitigation technology are:
a duty cycle limitation to less than 0.1%
a duty cycle limitation to less than 0.1% on one channel (300 kHz bandwidth)
a power reduction to less than 250 nW
Devices respecting this interference mitigation technology can coexist with devices using LBT.
Note: subject referred to RM for considerations as agreed at ERM#23 (see summary record) 
Supporting
Organizations
Biotronik ; BMWA ; Fed. ministry for Transport ; Fraunhofer Gesellschaft ; OFCOM 

 

Keywords Projects Clusters Frequencies Mandates Directives
  RADIO
SRD
TESTING
  Better Living with ICT
ULP-AMI (402 MHz to 405 MHz)
T-353
 
Official Journal
     
Remarks  

ETSI Home Page Any comments or problems with this application? Please let us know...